The AMI Watchdog
is a low cost state-of-the-art
trace oxygen analyzer designed for applications
requiring only essential features in
Class I, Div. 2 Groups C, D environments.
The Watchdog is well suited for a number
of applications including the natural
gas industry. It provides the essential
components required for a complete,
quick and simple installation where
trace oxygen analysis is required and
advanced features of the Model 2010BR
may become
redundant.
Click
Here For a Free Quote
|
Prior to the release of the
AMI Watchdog, competitive low cost solutions
consisted of either a two-wire transmitter
or an O2 sensor and cellblock. Both are
inadequate by themselves since they lack
an essential sample system and basic electronic
components. A trace oxygen analyzer is a
great leak detector; inadvertently adding
substandard valving and fittings after the
purchase of such a device can create minute
leak paths into the analyzer, which can
cause disastrously misleading results.
This is why AMI goes the extra
distance with our patented cellblock technology.
AMI's patented cellblock begins
with a single compact metallic block that
undergoes a complex machining process, followed
with corrosion-resistant nickel plating.
The block contains a series of drilled intersecting
gas passages and mounted directly on it
are: a custom designed flow control valve,
3-way selector valve (for sample/span/shut
off control), flow meter and 3 gas connection
fittings. All become an integral part of
the cellblock.
This eliminates the techniques
used by competitors of piecing several lengths
of tubing between numerous compression fittings,
valving and a flow meter. In using this
approach, they add many potential leak paths,
increase response times and the need for
additional space.
|
The Watchdog essentials include:
- Patented cellblock
technology
- 10 User-selectable
analog output ranges
- 2 User selectable
analog outputs:
1-5VDC or 4-20mA. selectable via available
software
- LCD
readout from 0.0ppm to 25.0%
- Advanced
calibration feature - calibrates
flow computer or PLC to the analyzers
analog
output allowing the PLC or flow computer
to track the analyzer precisely
- Remaining
oxygen sensor life indication
When the application requires, we can
provide one of our patent pending H2S
resistant oxygen sensors (up to 500ppm
H2S exposure for
its warranted life) to meet your needs.
This break-through sensor technology
eliminates the need for maintenance intensive
H2S scrubbers.
This rugged, compact, complete oxygen
analyzer allows for quick and simple installation
resulting in big savings.
 |
 |
|
AMI
guarantees higher quality
and exceptional services
at a lower price.
REQUEST
QUOTE
Or
call us at: 714.848.5533
|
|
 |
 |
|
|
|
|
|
|
|
- 10 user selectable output
ranges to choose from
- High resolution 3 digit
LCD
- RFI protected
- 1-5VDC and 4-20mA. isolated
analog output signals
- RS-232 bi-directional
communication for advanced
features
- Advanced analog output
calibration feature
- Power requirements: 10-28VDC
- Low minimum detection
limit
- Excellent repeatability
- Fast upscale/downscale
response times
- Patented Cellblock Technology:
Allows for all components
such as: flow control valve,
flow meter, 3-way calibration
valve, Sample/Span/Off and
compression fittings to
be an integral part of the
cellblock, eliminating long
lengths of tubing and fittings.
Other benefits of this design
include: compact size, faster
response times and front
panel sensor access without
the need for tools.
- Area Classification: Designed
to meet requirements for
Class 1, Div. 2, Groups
C,D application
- Unaffected by changes
in flow rate from 0.1 to
2.0 SCFH
- Wall mount or 2.0"
pipe
- Compact size
- 2 year warranty for analyzer,
parts and labor
- 6 month sensor warranty,
life expectancy 1-2 years
|

- 10
user selectable output
ranges to choose from:
0-10ppm, 0-50ppm, 0-100ppm,
0-500ppm, 0-1000ppm,
0-.500%, 0-1%, 0-5%,
0-10% and 25%
The user selection
of an output range simultaneously
controls the analog
output so that both
functions operate on
the same range
- Digital
display: High
resolution 3 digit LCD.
Reads full scale from
0.0ppm to 25.0% independently
of output range selection
- Isolated
analog output signal:
1-5VDC or 4-20mA
- Represents
the output range selected:
0-10ppm, 0-50ppm, 0-100ppm,
0-500ppm, 0-1000ppm,
0-.500%, 0-1%, 0-5%,
0-10% and 25%
- Power
requirements:
10-28VDC
<50mA. @ 12VDC
- Minimum
detection:
50ppb of oxygen
- Repeatability:
+/- 1% of range or +/-
0.2ppm of oxygen, whichever
is greater
- Operating
temperature range:
25° to 115°
F
- Diurnal
temperature specification:
< +/- 3 % of scale
over temperature range
- 90%
upscale response times:
10ppm - 25% <10 seconds
0-10ppm < 25 seconds
Typical downscale response:
1 minute exposure to
air down to 10ppm: <
15 minutes
- Area
Classification:
Designed to meet requirements
for Class 1, Div. 2,
Groups C,D applications
- Inlet
gas pressure:
0.5 to 150psig
- Gas
connections:
¼" 316 S.S.
compression fittings
- Wetted
parts: 316 S.S.
fittings, electroless
nickel plated cellblock,
gold plated contacts,
acrylic flow meter and
Viton O-rings
- Unaffected
by changes in flow rate
from 0.1 to 2.0 SCFH
- Mounting:
Wall mount or 2.0"
pipe
- Dimensions:
7.0"W x 6.5"H
x 4.5"D
- Weight:
6 lbs.
|
|
|
|
|
|
|
|
|
|
|
|
|