The development of the AMI
Model 2010BR trace oxygen analyzer
began with a collection of inputs from
hundreds of industry leading engineers
and field technicians in the natural
gas and petrochemical industries.
These inputs coupled with our advanced
engineering techniques have resulted
in a totally complete, compact, state-of-the-art
oxygen analyzer for harsh, hazardous
applications.
We have outfitted this unit with a
comprehensive state-of-the-art electronics
and complete sample handling package
using our patented cell block technology.
The front panel user interface is very
intuitive for ease of operation and
for the advanced users we provide an
RS-232 connection to access the more
advanced settings.
If necessary, this analyzer can be
mounted directly outside in all types
of weather. An optional high-tech, low
power proportional temperature controlled
heater allows for an expanded temperature
range of -20° to 115° F without
an additional enclosure or wiring. This
applies in either Class I, Div. 1 or
Div. 2 Groups B,C,D, area classifications,
allowing for a simple straightforward
installation.
Click
Here For a Free Quote
|
AMI's advanced microprocessor
based electronics platform solves all known
application requirements without adding
additional costly components.
Standard features include:
- Cutting
Edge Microprocessor Based Electronics
- High-resolution
single range 3 ½ Digit LCD
- 10
user-selectable output ranges:
0-10ppm, 0-50ppm, 0-100ppm, 0-500ppm,
0-1000ppm, 0-.500%, 0-1%, 0-5%, 0-10%
and an air calibration range of 0-25%
The user selection of an output range
simultaneously controls both alarm functions,
the analog output and the data logger
so that all 4 functions operate on the
same measurement range.
- 2
independent, fully adjustable alarm settings
with relay contact closures
- User
selectable alarm logic: latching
or non-latching, "Open on Alarm"
or "Close on Alarm"
- Programmable
alarm delays: Settings range from
0-300min. This feature prevents alarm
conditions due to quick O2
spikes or short-term operator error.
- Alarm
bypass button: Prevents false alarms
while performing monthly calibrations.
User settable from 0-120 minutes
- Data
logger: Logs data for 15 days @
1min intervals and 30 days @ 2min. intervals,
etc.
- Cal
Factor: indicates remaining O2
sensor life
- User
selectable security settings
- Isolated
1-5VDC and 4-20mA. analog outputs
- Simple
intuitive front panel user interface
- Available
in either AC or DC power
- Available
with or without explosion proof heater
AMI's patented cellblock begins
with a single compact metallic block that
undergoes a complex machining process, followed
with corrosion-resistant nickel plating.
The block contains a series of drilled intersecting
gas passages and mounted directly on it
are: a custom designed flow control valve,
3-way selector valve (for sample/span/shut
off control), flow meter and 3 gas connection
fittings. All become an integral part of
the cellblock.
|
This eliminates the techniques used by
competitors of piecing several lengths
of tubing between numerous compression
fittings, valving and a flow meter. In
using this approach, they add many potential
leak paths, increase response times, the
need for additional space and additional
employee labor.
Lastly, our front panel sensor access
makes the O2 sensor installation or replacement
quick and easy without tools or the need
to remove any sensitive components. It
also allows for a rapid, accurate calibration
on ambient air (20.9%) without the use
of tools. A span gas port comes standard
for those who prefer a certified span
gas calibration.
The accurate air calibration takes less
than 1 minute from start to finish!
Our advanced design approach has separated
us from the competition with the highest
performance, yet lowest cost oxygen analyzer
in the industry.
When the application
requires, we can provide one of our patent
pending H2S resistant
oxygen sensors (up to 500ppm H2S
exposure for its warranted life) to meet
your needs. This breakthrough sensor technology
eliminates the need for maintenance intensive
H2S scrubbers.
 |
 |
|
AMI
guarantees higher quality
and exceptional services
at a lower price.
REQUEST
QUOTE
Or
call us at: 714.848.5533
|
|
 |
 |
|
|
|
|
|
|
|
- 10 user selectable output
ranges to choose from
- High resolution 3 ½
digit LCD
- 2 fully adjustable oxygen
concentration alarms
- Alarm delays
- Alarm hold off/Bypass
- RFI protected
- 1-5VDC and 4-20mA. isolated
analog output signals
- Data logger
- RS-232 bi-directional
communication for advanced
features
- Advanced analog output
calibration feature
- Power requirements: Choice
of 10-28VDC or 115VAC power
- Low minimum detection
limit
- Excellent repeatability
- Extended operating temperature
range
- Fast upscale/downscale
response times
- Patented Cellblock Technology:
Allows for all components
such as: flow control, flow
meter, 3-way calibration
valve, Sample/Span/Off and
compression fittings to
be an integral part of the
cellblock, eliminating long
lengths of tubing and fittings.
Other benefits of this design
include: compact size, faster
response times and front
panel sensor access without
the need for tools.
- Area Classification: Approved
by CSA International to
UL and CSA standards as
meeting requirements for
Class1, Div. 1, Groups B,C,D
with a flammable gas sample
- Unaffected by changes
in flow rate from 0.1 to
2.0 SCFH
- Wall mount or 2.0"
pipe
- Compact size
- 2 year warranty for analyzer,
parts and labor
- 6 month sensor warranty,
life expectancy 1-2 years
|

- 10
user selectable output
ranges to choose from:
0-10ppm, 0-50ppm, 0-100ppm,
0-500ppm, 0-1000ppm,
0-.500%, 0-1%, 0-5%,
0-10% and 25%
The user selection
of an output range simultaneously
controls the two alarms,
the analog output and
the data logger so that
all 4 functions operate
on the same range
- Digital
display: High
resolution 3 ½
digit LCD. Reads full
scale from 0.00ppm to
25.0% independently
of output range selection
- Alarms:
2 fully adjustable oxygen
concentration alarms
Dry contacts 5A. @24VDC/115VAC
- Alarm
delays: Programmable
from 0-300 minutes
- Alarm
hold off: Programmable
from 0-120 minutes
- Isolated
analog output signal:
1-5VDC or 4-20mA
- Represents
the output range selected:
0-10ppm, 0-50ppm, 0-100ppm,
0-500ppm, 0-1000ppm,
0-.500%, 0-1%, 0-5%,
0-10% and 25%
- Data
logger: Logs
data for 15 days @ 1
minute intervals, 30
days @ 2 minute intervals,
etc.
Represents the output
range selected: 0-10ppm,
0-50ppm, 0-100ppm, 0-500ppm,
0-1000ppm, 0-.500%,
0-1%, 0-5%, 0-10% and
25%
- Power
requirements:
10-28VDC/ 115VAC
<70mA. @ 12VDC non
heated
<24W @12VDC with
heated option
- Minimum
detection:
50ppb of oxygen
- Repeatability:
+/- 1% of range or +/-
0.2ppm of oxygen, whichever
is greater
- Operating
temperature range:
25° to 115°
F non-heated
-20° to 115°
F heated option
- Diurnal
temperature specification:
< +/- 3 % of scale
over temperature range
- 90%
upscale response times:
10ppm - 25% <10 seconds
0-10ppm < 25 seconds
Typical downscale response:
1 minute exposure to
air down to 10ppm: <
15 minutes
- Area
Classification:
Approved by CSA International
to UL and CSA standards
as meeting requirements
for Class 1, Div. 1,
Groups B,C,D with a
flammable gas sample
- Inlet
gas pressure:
0.5 to 150psig
- Gas
connections:
¼" 316 S.S.
compression fittings
- Wetted
parts: 316 S.S.
fittings, electroless
nickel plated cellblock,
gold plated contacts,
acrylic flow meter and
Viton O-rings
- Unaffected
by changes in flow rate
from 0.1 to 2.0 SCFH
- Mounting:
Wall mount or 2.0"
pipe
- Dimensions:
13.0"W x 9.5"H
x 5.0"D
- Weight:
16 lbs.
|
|
|
|
|
|
|
|
|
|
|
|
|
|