The development of the AMI
Model 210BR began with a collection
of inputs from hundreds of industry
leading engineers and field technicians
in the natural gas, Process, and Petrochemical
business.
These inputs coupled with our advanced
engineering techniques has resulted
in a totally complete, compact, and
versatile package that can truly be
defined as state-of-the-art where oxygen
analysis is required in flammable gas
streams.
If necessary, the Model 210BR can be
mounted directly outside in all types
of weather. An optional high-tech, low
power proportional temperature controlled
heater allows for an expanded temperature
range of -40 to 115 F without an additional
enclosure or wiring. This applies in
either Class 1, Div. 1 or Div. 2 Groups
B,C,D, area classifications.
The rugged, compact, complete design
allows for quick and simple installation
resulting in big savings.
Click
Here For a Free Quote
|
AMI's advanced microprocessor
based electronics offers everything but
the kitchen sink, standard.
For example:
- High-resolution single
range LCD. Immediately reads from .01%
to 25% without range changing delays or
the need for manual adjustments.
- Microprocessor pulsed latching
alarm relays. The latching relay is pulsed
by the microprocessor controlling the
logic state to either open or closed.
This eliminates a continuous power drain
when using a solar panel and battery for
power.
- Programmable alarm delays:
Settings range from 0-300min. This feature
prevents alarm conditions due to quick
O2 spikes or short-term operator error.
- User selectable alarm setting:
latching, non-latching, fail safe non-failsafe
- 4-User selectable output
ranges: 0-1%, 0-5%, 0-10%, and 0-25%
- Alarm hold off button:
Prevents false alarms when routinely calibrating
and allows you to immediately silence
(for a user programmable period
of time of 0-300min.) an alarm condition
when necessary at the touch of a button.
- Data logger. Logs data
for 30 days @ 1min intervals.
- Complete security feature:
Front panel settings will not respond
to unwanted tampering if the security
feature has been initiated via. laptop
computer.
- Three standard isolated
outputs to chose from: 0-1VDC, 1-5VDC
and 4-20ma.
- Available in A/C
or DC power.
|
The
210BR incorporates our advanced patented
cellblock technology consisting of a single
compact metallic block that undergoes
a complex machining process, followed
with corrosion-resistant nickel plating.
The block contains a series of drilled
gas passages eliminating the need for
long lengths of tubing and compression
fittings.
A metering valve,
flow meter and a 3-way sample/span/shut
off valve are also an intregal part of
the patent cellblock. This sophisticated
design results in a quicker up and down
scale response times and eliminates potential
leak sources.
When the application
requires, we can provide a number of H2S
resistant oxygen sensors to meet your
needs.
 |
 |
|
AMI
guarantees
higher quality and exceptional
services
at a lower price.
REQUEST
QUOTE
Or
call us at:
714.848.5533
|
|
 |
 |
The front panel sensor
access makes sensor installation or replacement
quick and easy. It also allows for a rapid,
accurate calibration on ambient air (20.9%)
without the use of tools. A span gas port
comes standard for those who prefer a
certified span gas calibration.
Our advanced design approach has separated
us from the competition with the highest
performance, yet lowest cost trace oxygen
analyzer in the industry.
|
|
|
|
|
|
-
Standard
user selectable output
ranges:
0-1%, 0-5%, 0-10%, and
0-25%
-
Sensitivity:
0.5% Full scale
-
Repeatability:
±1% of full scale
at constant temperature
-
Display:
High resolution single
range 3½ digit
LCD
-
Operating
temperature range:
32-115°F non heated
-40-115 °F with optional
heater
-
Diurnal
temperature specification:
32-115°F, < ±3
of scale
-40-115 °F with optional
heater < ±3
of scale
-
90%
Upscale/downscale Response
times:
0-25% less than 15 sec.
for P-2 sensor
0-25% less than 30 sec.
for P-3 sensor
-
Alarms
(standard):
2 Fully adjustable oxygen
concentration alarm set
points. Dry contacts 5A.
-
Isolated
output signals:
1-5VDC, and 4-20ma.
-
Power
requirements:
10-28VDC or 115VAC<
40mA (non heated). 30
Watts approx. with heated
option
-
Mounting:
Wall mount or 2"
pipe
-
Area
classification:
Designed for use in Class
I, Div. I, Groups BCD
-
Dimensions:
10.00W x 13.00H x 5.00D
-
Weight:
13 lbs
-
Sensor
access:
Easy, direct front panel
access
-
Sample
connections:
316 Stainless steel ¼"
Compression
-
Wetted
parts:
Nickel-plated metallic
|

- Designed to meet UL-913
and UL-1203
- Class I, Div. I.,
Groups BCD with flammable
sample. Approval pending.
- High resolution single
range 3 ½ LCD
- 4-User selectable
output ranges
- 2 Fully adjustable
oxygen concentration
alarms. Form C relay
contact closures
- Programmable alarm
delays. 0-300min.
- Alarm hold off: Programmable
0-120.
- Data logger
- RS232 bi-directional
communications
- Isolated outputs:
1-5VDC, or -20ma.
- Power requirements:
10-28VDC or 115VAC
< 40mA (non heated).
30 Watts approx. with
heated option
- Direct, easy front-panel
sensor access
- Panel mount or 2"
pipe mount
- Compact size
- Nickel-plated cellblock,
316 stainless steel
fittings
- 2 Year warranty (excluding
sensor)
- No calibration or
support gases necessary
- Virtually unaffected
by hydrocarbons and
other oxidizable gases
- RFI protected
- Unaffected by change
in flow rate (0.1-5
SCFH)
|
|
|
|
|
|
|
|
|
|
|
|
|