The development of the AMI
Model 2010BR began with a collection
of inputs from hundreds of industry
leading engineers and field technicians
in the natural gas and Petrochemical
business.
These inputs coupled with our advanced
engineering techniques have resulted
in a totally complete, compact, and
versatile package that can truly be
defined as state-of-the-art where oxygen
analysis is required in flammable gas
streams.
If necessary, the Model 2010BR can
be mounted directly outside in all types
of weather. An optional high-tech, low
power proportional temperature controlled
heater allows for an expanded temperature
range of 40 to 115°F without
an additional enclosure or wiring. This
applies in either Class 1, Div. 1 or
Div. 2 Groups B,C,D, area classifications.
The rugged, compact, complete design
allows for quick and simple installation
resulting in big savings.
|
AMIs advanced microprocessor
based electronics offers everything but
the kitchen sink, standard. For example:
-
High-resolution
single range LCD immediately reads from
.01ppm to 25% without range changing
delays or the need for manual adjustments.
-
Microprocessor
pulsed latching alarm relays. The latching
relay is pulsed by the microprocessor
controlling the logic state to either
open or closed. This eliminates a continuous
power drain when using a solar panel
and battery for power.
- Programmable alarm delays:
Settings range from 0-300min. This feature
prevents alarm conditions due to quick
O2 spikes or short-term operator error.
-
User selectable
alarm setting: latching, non-latching,
fail safe non-failsafe
-
10-user
selectable output ranges: 0-10ppm, 0-50ppm,
0- 100ppm, 0-500ppm, 0-1000ppm, 0-.500%,
0-1%, 0-5%, 0- 10% and an air calibration
range of 0-25%
-
Alarm
hold off button: Prevents false alarms
when routinely calibrating and allows
you to immediately silence (for a user
programmable period of time of 0-300min.)
an alarm condition when necessary at
the touch of a button.
-
Data logger.
Logs data for 30 days at 1min intervals.
-
Complete
security feature: Front panel settings
will not respond to unwanted tampering
if the security feature has been initiated
via laptop computer.
-
Three
standard isolated outputs to chose from:
0-1VDC, 1-5VDC or 4-20mA.
-
Available
in A/C or DC power.
The 2010BR incorporates our
advanced patented cellblock technology consisting
of a single compact |
metallic block that
undergoes a complex machining process,
followed with corrosion-resistant nickel
plating. The block contains a series of
drilled gas passages eliminating the need
for long lengths of tubing and compression
fittings; it uses critical orifices in
place of needle valves and pressure regulators;
it also contains a unique integrated 3-way
sample/span/shut off valve, and flowmeter.
This sophisticated design results in quicker
up and down scale response times and eliminates
potential leak sources.
When the application requires, we add
a quick-change H2S scrubber and particulate/coalescing
filter mounted directly to the block.
Contents of the H2S scrubber can be replenished
in minutes without the use of any tools
or potting compound.
The front panel sensor access makes sensor
installation or replacement quick and
easy. It also allows for a rapid, accurate
calibration on ambient air (20.9%) without
the use of tools. A span gas port comes
standard for those who prefer a certified
span gas calibration.
Our advanced design approach has separated
us from the competition with the highest
performance, yet lowest cost trace oxygen
analyzer in the industry.
|
|
|
|
|
|
-
10-standard
user selectable output
ranges:
0-10ppm
0-50ppm
0-100ppm
0-500ppm
0-1000ppm
0-.500%
0-1%
0-5%
0-10%
and a calibration range
of 0-25%
-
Minimum
detection :
0.5% of range or 50ppb
-
Repeatability:
±1% of full scale
at constant temperature
(on 10ppm output range
the repeatablility is
±3% of scale)
-
Display:
High resolution single
range 3½ digit
LCD
-
Operating
temperature range:
32-115°F non heated
-40-115 °F with optional
heater
-
Diurnal
temperature specification:
32-115°F,
< ±3 of scale
-40-115 °F with optional
heater
< ±3 of scale
|
-
90%
Upscale response
times:
100ppm-25%
< 10 sec.
0-10ppm < 25
sec.
0-5ppm < 50 sec.
-
Alarms
(standard):
2 Fully adjustable
oxygen concentration
alarm set points.
Relay contacts 5A/115VAC
-
Isolated
output signal:
0-1VDC, 1-5VDC,
or 4-20mA
-
Power
requirements:
10-28VDC
or 115VAC <
40mA
(non heated)
30 Watts approx.
with heated option
-
Mounting:
Wall mount or
2" pipe
-
Area
classification:
Designed for use in
Class I, Div. I, Groups
BCD
-
Dimensions:
10.00"
W x 16.75" H
x 6.00" D
-
Weight:
15 lbs
-
Sensor
access:
Easy, direct front
panel access
-
Sample
connections:
316 S.S. ¼"
compression fittings
-
Wetted
parts:
Nickel-plated metallic

- Designed to meet UL-913
and UL-1203
- Class I, Div. I.,
Groups B, C, D with
flammable sample.
|
- High resolution single
range 3½ LCD
- 10-user selectable output
ranges
- 2 fully adjustable oxygen
concentration alarms. Programmable
relay contact closure logic.
- User programmable alarm
delays. 0-300min.
- Alarm hold off: Programmable
0-300min.
- Complete system security
- Data logger
- RS232 bi-directional communications
- Isolated outputs: 0-1VDC,
1-5VDC, or 4-20mA
- Power requirements:
10-28VDC or 115VAC <
40mA (non heated)
30 Watts approx. with heated
option
- Direct, easy
front-panel sensor access
- Panel mount
or 2" pipe mount
- Compact size
- Nickel-plated
cellblock, 316 S. S. compression
fittings
- 2 year warranty
(excluding sensor)
- No calibration
or support gases necessary
- Virtually
unaffected by hydrocarbons
and other oxidizable gases
- RFI protected
- Unaffected
by change in flow rate (0.1-5
SCFH)
|
|
|
|
|
|
|
|
|
|
|
|
|