The AMI Model
2010B is a complete package for accurately
and continuously measuring trace levels
of oxygen in flammable gas streams.
The sophisticated microprocessor
based electronics provides advanced features
with a simple straightforward user interface.
The controls are designed to be intuitive,
quick and simple while still providing a
security function against unauthorized tampering.
The large 3½ digit
display auto-ranges for maximum resolution,
while the user selectable output range remains
fixed in your range of interest. The 2010B
is supplied with ten standard output ranges:
0-10ppm, 0-50ppm, 0-100ppm, 0-500ppm, 0-1000ppm,
0-.500%, 0-1%, 0-5%, 0-10% and a air calibration
range of 0-25%, two fully adjustable oxygen
alarm settings, and a 1-5VDC output.
|
The RS-232 port allows advanced users
to quickly and easily take advantage of
all the standard features:
-
Individual
alarm delay setting
(0-300 minutes)
-
Choice
of latching/non-latching alarms
-
Failsafe
or non-failsafe logic
-
Alarm pulse
feature: Operates a low powered latching
slam valve in remote solar powered locations
-
Output
range default: Automatically returns
to your output range of interest in
case of technician error or unwanted
tampering.
The 2010B incorporates our advanced patented
cellblock technology consisting of a single
compact metallic block that undergoes
a complex machining process, followed
with corrosion-resistant nickel plating.
The block contains a series of drilled
gas passages eliminating the need for
long lengths of tubing and compression
fittings; it uses critical orificesin
place of needle valves and pressure regulators;
it also contains a unique integrated 3-way
sample/span/shut off valve, and flowmeter.
This sophisticated design results in quicker
up and down scale response times and eliminates
potential leak sources.
When the application requires,
we add a quick-change H2S scrubber and particulate/coalescing
filter mounted directly to the block. Contents
of the H2S scrubber can be replenished in
minutes without the use of any tools or
potting compound.
|

The front panel sensor access makes sensor
installation or replacement quick and
easy. It also allows for a rapid, accurate
calibration on ambient air (20.9%) without
the use of tools. A span gas port comes
standard for those who prefer a certified
span gas calibration.
The 2010B does not require
a zero or other support gases to operate
and is linear throughout its wide range
of measurement.
The Model 2010B has been agency
approved for Class I, Div. II, Groups C,
D in a flammable gas sample to UL standard
913.
|
|
|
|
|
|
-
Standard
user selectable output
ranges:
0-5ppm, 0-10ppm, 0-50ppm,
100ppm, 0-500ppm,, 0-1000ppm,
0-.500%, 0-1%, 0-5%,
0-10% and a calibration
range of 0-25%
-
Minimum
detection:
0.5% of range or 50ppb
-
Repeatability:
±1% of full scale
at constant temperature
(for 10ppm, repeatability
is ±3% of scale)
-
-
Operating
temperature range:
32°-113°F
(0-45°C)
-
Diurnal
temperature specification:
0-45°C,
< ±3 of scale
-
90%
Upscale response
times:
100ppm-25% <
10 sec
0-10ppm < 25
sec
0-5ppm < 50 sec
-
Alarms
(standard):
2 fully adjustable oxygen
concentration set points.
Solid state contact
closure. 1A 5-28VDC
-
Trouble
contacts:
Solid state contact
closure. 1A/5-28VDC
-
Output
signal:
1-5VDC, or optional
isolated 4-20mA
-
Power
requirements:
10-28VDC <40mA
-
Sample
connections:
316 S.S. ¼"
compression
fittings
|
-
Major
Wetted
parts:
Nickel-plated
metallic
block,
316
S.S.
fittings
-
-
Area
classification:
Approved for Class
1, Div. 2, Groups
C,D in a flammable
gas sample to UL Standard
913-88.
-
Dimensions:
10.00"W x 16.75"H
x 6.00"D as shown
-
-
Sensor
access:
Easy, direct front
panel access
- NRTL Approval to UL-913
Class I, Div. II, Groups
CD with flammable sample
- Auto-ranging digital display
- 11 User selectable output
ranges: 0-5ppm, 0-10ppm,
0-50ppm 100ppm, 0-500ppm,
0-1000ppm, 0-5000ppm, 0-1%,
0-5%, 0-10% and a calibration
range of 0-25%
- Programmable alarm delays
(0-300min.)
- Output range default (0-188min.)
- Delay for alarm set points
(0-300min.)
- RS232 bi-directional communications
- Output: 1-5VDC, or isolated
4-20mA
- Power requirements: 10-28VDC
< 40mA
- 2 Fully adjustable oxygen
concentration alarms. Solid
state "contact closures"
- Trouble alarm. Solid state
"contact closures"
|
- Security
function
- Direct, easy front-panel
sensor access
- Compact size
- Nickel-plated
cellblock, 316 stainless
steel fittings
- 2 Year warranty
(excluding sensor)
- No calibration
or support gases necessary
- Virtually
unaffected by hydrocarbons
and other oxidizable gases
- RFI protected
- Unaffected
by change in flow rate
(0.1-5 SCFH)
-
Revolutionary
cellblock technology
-
Faster
response and recovery
times
-
Direct
front panel sensor
access
-
-
(11)
customer selectable
output ranges
-
Programmable
alarm delays (0-188min.)
-
Output
range default setting
(0-188min.)
-
Built
in intrinsically safe
barriers for input
power and alarm connections
-
-
Unaffected
by change in flow
between
0.1-5 SCFH
-
No
support gases necessary
-
Unaffected
by hydrocarbon or
other oxidizable gases
-
-
-
NRTL
Approval to UL-913
Class I, Div. II,
Groups CD with flammable
sample
|
|
|
|
|
|
|
|
|
|
|
|